VIG - microprocessor based hydrocarbon analyzer manufacturer
Flame Ionization Detector Based Hydrocarbon Analyzers
Environmental Instrumentation Manufacture, Sales, Service & Rent

VIG - air quality monitoring equipment manufacturer

Home | Products | CEMS | Application Form | News | Sitemap | Sales and Service | Literature Request | Rentals | Clients | Links | Contact Us

Model 210 Hydrocarbon Analyzer -
 FID Oven Heated Methane / Ethane / Non Methane Ethane (
NMEHC) / Total  Hydrocarbon Analyzer

Model 210 Hydrocarbon Analyzer

The VIG Industries, Inc. Model 210 is a microprocessor based, oven heated  methane / ethane /  non methane-ethane (NMEHC) total hydrocarbon analyzer for high accuracy, sensitivity and stability. The Model 210 measures methane, ethane, non methane-ethane hydrocarbons at preset intervals and continuously measures total hydrocarbons in real time. The Model 210 uses two (2) Flame Ionization Detectors (FID).  One FID is used to obtain methane, ethane and non methane ethane hydrocarbon readings. The second FID is used to continuously measure total hydrocarbons.  A sample is fed to the analyzer via an internal heated pump. After the sample reaches the analyzer it is fed to the first FID to measure total hydrocarbons. A small portion of the sample is trapped and fed through a GC (gas chromatography) column to separate the methane and ethane from the total and is fed to the second FID detector for analysis. As soon as the methane and ethane components are read, the remaining sample is back flushed through the column and fed to the FID detector for analysis of  the non methane ethane component. The LCD display indicates the amount of methane, ethane, residual (non methane ethane) and total hydrocarbons.

The Model 210 utilizes two flame ionization detectors (FID). The detectors, all sample lines, GC column and all solenoid valves that come in contact with the sample are maintained in a controlled, heated oven to prevent condensation, and to provide reliable performance in the analysis of concentrations of wide varieties of hydrocarbons in gaseous mixtures and in ambient air.

Model 210 Hydrocarbon Analyzer PDF (1MB)Download 1MB

Applications

  • Compliance Monitoring - U.S. E.P.A. Method 18 and Method 25A
  • Process Monitoring - Continuous monitoring and alarm or control of:
    • Process gas streams utilizing organic solvents
    • Crude oil
    • Other chemicals containing hydrocarbons
  • Efficiency Monitoring - Monitoring effluent of volatile organic compound (VOC) reduction equipment for:
    • Environmental compliance
    • Efficiency control of incinerators (Thermal or catalytic)
    • Scrubbers
    • Carbon absorbers
    • Monitoring of catalytic converters
    • Other abatement equipment
  • Safety Monitoring - Lower explosive limit (LEL) monitoring and/or control of:
    • Ovens/dryers
    • Fugitive emissions monitoring
    • Personnel work area monitoring
    • Leak detection of process equipment or solvent storage areas
  • Stack Monitoring

Features

  • Easy to use software
  • GC technology used for separation
  • Automatic ignition
  • Automatic fuel shut-off system
  • Automatic flame-out indicator
  • Adjustable alarm and oven settings
  • Precision 1% of full scale
  • Fast response
    • Methane, Ethane and Non methane - minimum 3.0 minute cycle
    • Total Hydrocarbons - Within 5 seconds to 90% of final reading
      (continuous real time reading)
  • Two stage sample filter with exchangeable stainless steel elements
  • PTFE isolated detector
  • 19" rack or bench mount
  • Heated Sample pump
  • Outputs 0-1, 0-5, or 0-10Volts

Options

  • Zero and calibration input ports with software
  • 4-20mA output - no extra charge
  • Range recognition relays
  • RS-232 interface
  • Internal combustion air supply
  • Automatic sample shut-off system

Related Available Equipment

  • External zero air generator (reduces bottles)
  • Hydrogen Generator (reduces bottles)
  • Heated sample lines and controllers
  • Strip chart recorders and data loggers
  • NEMA rated enclosures

Standard Specifications

 

Analysis Method

Flame Ionization Detector (FID)

Separation Method

Column

Precision

+/- 1% of Full Scale

Response Time

  • Total Hydrocarbons - Within 5 seconds to 90% of final reading
    (continuous real time reading)
  • Methane - Approximately 40 seconds,
    updated every 3 minutes
  • Ethane - Approximately 55 seconds,
    updated every 3 minutes
  • Non methane - Approximately 90 seconds,
    updated every 3 minutes

Zero Drift

+/- 1% of Full Scale Per 24 Hours

Span Drift

+/- 1% of Full Scale Per 24 Hours

Linearity

Within 1%

Oxygen Synergism

Within 1% of Selected Range

Ranges

Standard Ranges (4 Ranges per amplifier, two amplifiers per analyzer,
one amplifier for total hydrocarbons,
one amplifier for methane, ethane and non methane-ethane)
  • 0-100, 1000, 10000, 100000ppm (LDL 0.1ppm)
  • 0-10, 100, 1000, 10000ppm (LDL 0.01ppm)

 

Display

Backlit Digital / Graphical LCD

Outputs

0-1, 5, or 10 Volts and 4-20mA.

Alarms

Flame Out & Concentration, Latching

Safety

Flame Out Indicator with Fuel Shut-Off System

Oven Temperature

Operating Temperature 275F, Adjustable 200 to 300F

Sample Pump

Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional

Ignition Automatic (can be set to manual by operator)
Glow Plug Main and spare glow plug, selectable by switch on back panel
Dimensions 19" front panel, 16.75" wide chassis, 18" deep chassis,
21" deep with fittings and handles, 9" high
Weight 55 to 65 pounds depending on options

Operating Requirements

 
Fuel Hydrogen (UHP Grade) @ 18psi incoming pressure
Combustion Air Zero Grade Air @ 18PSI
Operation Air Compressed Air (Instrument Air - Dry / Oil-Less) @ 60PSI
Zero Gas Zero Air (UHP Grade) @ 10PSI
Calibration Gas User Selected @ 10PSI

Carrier Gas

Nitrogen (UHP Grade) @ 30PSI
Power 115Vac, 60Hz or optional 220Vac, 50Hz
Ambient Temperature 32 to 110 F

Warranty

All instruments sold by VIG Industries, Inc., are warranted for a period of one (1) year from the date of purchase, against defects in materials and workmanship. VIG warrants that the product supplied conforms to the specifications assigned thereto. There is no other warranty either implied or expressed. Seller liability is limited specifically to the cost or assigned value of the items sold.